S3030F

Single-Channel High voltage Precision Source Meter

The Semight S3030F is a compact and cost-effective single channel high voltage, high-power power supply/measurement unit that can simultaneously output and measure voltage and current, it can provide a maximum power output of ± 3500V, ± 120mA (DC), 10 G Ω, and 180W, and can be widely used in testing and research fields such as power semiconductor characteristics, GaN, SiC characterization, composite materials, and high-voltage leakage current. S3030F supports traditional SMU SCPI commands, making migration of test code easy and fast, it can support multi machine synchronization and be integrated into production testing systems to improve system testing efficiency and reduce costs.

  • High range – Range:±3500 V、±120 mA(DC)、180 W
  • High resolution – The minimum measurement resolution can reach 10 fA/100 μV
  • High sampling rate – Supports up to 1M ADC sampling rate
  • Threshold trigger – Hardware high-speed IO, capable of threshold triggering, enabling efficient interaction between output measurement values and user systems

Features and Benefits

DC I-V output capability

working conditions:
Temperature 23 °C ± 5 °C
Humidity 30% to 70% relative humidity
Measure after preheating for 60 minutes, and the ambient temperature change during measurement is less than ± 3 ° C
Calibration cycle 1 year
Measurement speed 1 PLC

Voltage Source specifications

Voltage setting accuracyRangeProgramming resolutionAccuracy (1 year)

± (% reading+offset)

Typical noise(RMS)

0.1Hz-10Hz

±3500 V40 mV0.02%+600 mV50 mV
±2500 V30 mV0.02%+450 mV40 mV
±1500 V20 mV0.02%+300 mV25 mV
±600 V7 mV0.02%+120 mV10 mV
±200 V3 mV0.02%+40 mV3 mV
Temperature Coefficient±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃)
Settling time<5 ms (typical)
Overshoot<± 1 % (Typical. Norma mode. Step is 10 % to 90 % range, full range, resistive load)
Noise 10Hz-20MHz1500 V voltage source, 120 mA resistive load, < 200 mV RMS

Current Source specifications

Current setting accuracyRangeProgramming resolutionAccuracy (1 year)

± (% reading+offest)

Typical noise (RMS)

0.1Hz-10Hz

±120 mA13 μA0.02% + 35 μA120 uA
±20 mA400 nA0.02% + 15 μA20 uA
±10 mA200 nA0.02% + 3 uA10 uA
±1 mA20 nA0.02% + 300 nA1 uA
±100 μA2 nA0.02% + 30 nA100 nA
±10 μA200 pA0.03% + 5 nA10 nA
±1 μA20 pA0.03% + 1 nA1 nA
±100 nA2 pA0.2% +100 pA+Vo x 100 fA100 pA
±10 nA200 fA0.2% + 10 pA+ Vo x 10 fA10 pA
±1 nA20 fA0.2% + 5 pA+ Vo x 1 fA1 pA
Temperature Coefficient±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃)
Settling time<10 ms (typical)
Overshoot<± 1 % (Typical. Norma mode. Step is 10 % to 90 % range, full range, resistive

load)

 

  1. 120mA range only supports 1500V and below voltage range

Voltage measurement specifications

Voltage measurement accuracy

 

 

 

 

 

RangeMeasurement resolutionAccuracy (1 year)

± (% reading+offset)

±3500 V1 mV0.02%+600 mV
±2500 V1 mV0.02%+450 mV
±1500 V1 mV0.02%+300 mV
±600 V100 μV0.02%+120 mV
±200 V100 μV0.02%+40 mV
Temperature Coefficient±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃)

Current measurement specifications

Current measurement accuracy

 

 

 

 

 

 

 

 

 

RangeMeasurement resolutionAccuracy (1 year)

± (% reading+oddset)

±120 mA1100 nA0.02% + 35 μA
±20 mA10 nA0.02% + 15 μA
±10 mA10 nA0.02% + 3 μA
±1 mA1 nA0.02% + 300 nA
±100 μA100 pA0.02% + 30 nA
±10 μA10 pA0.03% + 5 nA
±1 μA1 pA0.03% + 1 nA
±100 nA100 fA0.2% +100 pA+Vo x 100 fA
±10 nA10 fA0.2% + 10 pA+ Vo x 10 fA
±1 nA1 fA0.2% + 5 pA+ Vo x 1 fA
Temperature

Coefficient

±(0.15 × accuracy index)/°C (0℃-18℃,28℃-50℃)

 

  1. 120mA range only supports 1500V and below voltage range.