S0342C

S0342C

4-channel PXIe SMU

The S0342C is a high precision, compact, cost-effective, 4-channel PXIe Source/Measure Unit (SMU) with the capability to source and measure both voltage and current. It has maximum ±30 V, ± 500 mA DC/Pulsed sourcing capability, and supports conventional SMU SCPI commands for easy programming. It supports most standard PXIe chassis and multi-card synchronization. The S0342C can be integrated into production test environment to achieve test efficiency improvement and cost reduction.

Features

10

Easy to Expand

4-channel PXIe SMU, easy to scale out

11

High Range

Range: ± 30 V, ± 500 mA (DC/pulse)

12

High Resolution

The minimum measurement resolution can reach 100 pA/60 μV

13

High Sampling Rate

Up to 500K ADC sampling rate

Functions and Advantages

5 Functions In One Card

Voltage source
Current source
Ammeter
Voltmeter
Electronic load

5 Functions In One Card
Four Quadrants

♦ The first and third quadrants are the source: the actual polarity of output V / I follows the source setting.
♦ The second and fourth quadrants are for load: CC and CV cooperate. When the load is used, the polarity of the load setting is opposite to the source polarity.

Can Test Various Equipment

Can Test Various Equipment
Measurement Data

Capture More Measurement Data

♦ 6.5-digit resolution: Enjoy best-in-class 6.5-digit sourcing and measurement resolution.
♦ 100pA / 60μV resolution: excellent measurement sensitivity.
♦ 500K points / second sampling rate: provide high-speed measurement, and can quickly set / digitize the rate to any waveform generator (list scan).

Rich Scanning Function

Rich Scanning Function

Voltage Source Specifications

Voltage Programming Accuracy
Range Programming Resolution Accuracy (1 Year)
± (% reading+ offset)[1]
Typical Noise (RMS)
0.1 Hz-10Hz
±30 V[3] 1mV 0.03%+4 mV 1 mV
±6 V 200 μV 0.03%+1 mV 100 μV
Temperature Coefficient ±(0.15 × accuracy)/°C (0℃-18℃, 28℃-50℃)
Channel[2] CH0 to CH3
Output Power 3W per channel and 6W total
Settling Time <200us (typical)
Overshoot <±0.1% (typical. normal. step is 10% to 90% range, full range, resistive load)
Noise 10Hz-20MHz 6V voltage source, 0.5A resistive load, <3 mVrms

[1] Accuracy calculation example: To test the accuracy of a 6V range with a 1V output, the tolerance is:084fcd02c92a2e84700df2ffe5eb4179
[2] Channels are isolated from earth ground but share a common LO
[3] This instrument has potentially dangerous high voltage (±31.5 V) output to the HI / Sense HI / Guard terminals. To prevent electric shock, appropriate safety precautions must be taken before powering on. Do not connect the Guard terminal to any output, including shorting it to the chassis ground or output LO, otherwise the instrument will be damaged

Current Source Specifications

Current Programming Accuracy
Range Programming Resolution Accuracy (1 Year)
± (% reading+ offset)
Typical Noise (RMS)
1 Hz-10Hz
±500 mA[4] 20 μA 0.05% + 100 μA + Vo*25 μA 10 μA
±100 mA 4 μA 0.05% + 10 μA+ Vo*5 μA 1 μA
±10 mA 400 nA 0.05% + 5 μA + Vo*500 nA 100 nA
±1 mA 40 nA 0.05% + 500 nA + Vo*50 nA 10 nA
±100 μA 4 nA 0.05% + 50 nA + Vo*5 nA 1 nA
±10 μA 400 pA 0.05% + 20 nA + Vo*500 pA 150 pA
Temperature Coefficient ±(0.15 × accuracy)/°C (0℃-18℃, 28℃-50℃)
Channel CH0 to CH3
Output Power 3 W per channel and 6 W total
Settling Time <300 μs (typical)
Overshoot <±0.1% (typical. normal. step is 10% to 90% range, full range, resistive load)

[4] 500mA range is available only for 6V voltage range

Voltage Measurement Specifications

Voltage Measurement Accuracy
Range Measurement Resolution Accuracy (1 Year)
± (% reading+ offset)
±30 V 300 μV 0.03%+4 mV
±6 V 60 μV 0.03%+1 mV
Temperature Coefficient ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃)

Current Measurement Specifications

Current Measurement Accuracy
Range Measurement Resolution Accuracy (1 Year)
± (% reading+ offset)
±500 mA 10 μA 0.05% + 100 μA + Vo*25 μA
±100 mA 1 μA 0.05% + 10 μA+ Vo*5 μA
±10 mA 100 nA 0.05% + 5 μA + Vo*500 nA
±1 mA 10 nA 0.05% + 500 nA + Vo*50 nA
±100 μA 1 nA 0.05% + 50 nA + Vo*5 nA
±10 μA 100 pA 0.05% + 20 nA + Vo*500 pA
Temperature Coefficient ±(0.15 × accuracy)/°C (0℃-18℃, 28℃-50℃)

I-V Out Capability

I-V Out Capability-2

Typical Output Settling Time

Source Range Output Settling Time[5] Condition
Fast[6] Normal Slow
Voltage 30 V <500 μs <1.5 ms <3 ms Time required to reach within 0.1 % of final value at open load condition.
Step is 10 % to 90 % range
6 V <250 μs <780 μs <2.8 ms
Current 500 mA <150 μs <400 μs <2.5 ms Time required to reach within 0.1 % of final value at short condition.
Step is 10 % to 90 % range
100 mA~1 mA <120 μs <300 μs <2.5 ms
±100 μA <150 μs <1.5 ms <2.5 ms
±10 μA <700 μs <2 ms <2.5 ms

[5] Output transition rate: Three modes of Fast, Normal, and Slow are available, allowing users to adjust according to load characteristics to achieve appropriate setup time or stability. This feature only supports switching the output transition rate when OUTPUT is set to OFF
[6] Slow mode is recommended for overshoot sensitive equipment. Fast mode may have overshoot on output in some condition