S0342C
S0342C
4-channel PXIe SMU
The S0342C is a high precision, compact, cost-effective, 4-channel PXIe Source/Measure Unit (SMU) with the capability to source and measure both voltage and current. It has maximum ±30 V, ± 500 mA DC/Pulsed sourcing capability, and supports conventional SMU SCPI commands for easy programming. It supports most standard PXIe chassis and multi-card synchronization. The S0342C can be integrated into production test environment to achieve test efficiency improvement and cost reduction.
Features

Easy to Expand
4-channel PXIe SMU, easy to scale out

High Range
Range: ± 30 V, ± 500 mA (DC/pulse)

High Resolution
The minimum measurement resolution can reach 100 pA/60 μV

High Sampling Rate
Up to 500K ADC sampling rate
Functions and Advantages
5 Functions In One Card
Voltage source
Current source
Ammeter
Voltmeter
Electronic load


♦ The first and third quadrants are the source: the actual polarity of output V / I follows the source setting.
♦ The second and fourth quadrants are for load: CC and CV cooperate. When the load is used, the polarity of the load setting is opposite to the source polarity.
Can Test Various Equipment


Capture More Measurement Data
♦ 6.5-digit resolution: Enjoy best-in-class 6.5-digit sourcing and measurement resolution.
♦ 100pA / 60μV resolution: excellent measurement sensitivity.
♦ 500K points / second sampling rate: provide high-speed measurement, and can quickly set / digitize the rate to any waveform generator (list scan).
Rich Scanning Function

Voltage Source Specifications
| Voltage Programming Accuracy | ||||
| Range | Programming Resolution | Accuracy (1 Year) ± (% reading+ offset)[1] |
Typical Noise (RMS) 0.1 Hz-10Hz |
|
| ±30 V[3] | 1mV | 0.03%+4 mV | 1 mV | |
| ±6 V | 200 μV | 0.03%+1 mV | 100 μV | |
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃, 28℃-50℃) | |||
| Channel[2] | CH0 to CH3 | |||
| Output Power | 3W per channel and 6W total | |||
| Settling Time | <200us (typical) | |||
| Overshoot | <±0.1% (typical. normal. step is 10% to 90% range, full range, resistive load) | |||
| Noise 10Hz-20MHz | 6V voltage source, 0.5A resistive load, <3 mVrms | |||
[1] Accuracy calculation example: To test the accuracy of a 6V range with a 1V output, the tolerance is:
[2] Channels are isolated from earth ground but share a common LO
[3] This instrument has potentially dangerous high voltage (±31.5 V) output to the HI / Sense HI / Guard terminals. To prevent electric shock, appropriate safety precautions must be taken before powering on. Do not connect the Guard terminal to any output, including shorting it to the chassis ground or output LO, otherwise the instrument will be damaged
Current Source Specifications
| Current Programming Accuracy | ||||
| Range | Programming Resolution | Accuracy (1 Year) ± (% reading+ offset) |
Typical Noise (RMS) 1 Hz-10Hz |
|
| ±500 mA[4] | 20 μA | 0.05% + 100 μA + Vo*25 μA | 10 μA | |
| ±100 mA | 4 μA | 0.05% + 10 μA+ Vo*5 μA | 1 μA | |
| ±10 mA | 400 nA | 0.05% + 5 μA + Vo*500 nA | 100 nA | |
| ±1 mA | 40 nA | 0.05% + 500 nA + Vo*50 nA | 10 nA | |
| ±100 μA | 4 nA | 0.05% + 50 nA + Vo*5 nA | 1 nA | |
| ±10 μA | 400 pA | 0.05% + 20 nA + Vo*500 pA | 150 pA | |
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃, 28℃-50℃) | |||
| Channel | CH0 to CH3 | |||
| Output Power | 3 W per channel and 6 W total | |||
| Settling Time | <300 μs (typical) | |||
| Overshoot | <±0.1% (typical. normal. step is 10% to 90% range, full range, resistive load) | |||
[4] 500mA range is available only for 6V voltage range
Voltage Measurement Specifications
| Voltage Measurement Accuracy | ||||
| Range | Measurement Resolution | Accuracy (1 Year) ± (% reading+ offset) |
||
| ±30 V | 300 μV | 0.03%+4 mV | ||
| ±6 V | 60 μV | 0.03%+1 mV | ||
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃) | |||
Current Measurement Specifications
| Current Measurement Accuracy | ||||
| Range | Measurement Resolution | Accuracy (1 Year) ± (% reading+ offset) |
||
| ±500 mA | 10 μA | 0.05% + 100 μA + Vo*25 μA | ||
| ±100 mA | 1 μA | 0.05% + 10 μA+ Vo*5 μA | ||
| ±10 mA | 100 nA | 0.05% + 5 μA + Vo*500 nA | ||
| ±1 mA | 10 nA | 0.05% + 500 nA + Vo*50 nA | ||
| ±100 μA | 1 nA | 0.05% + 50 nA + Vo*5 nA | ||
| ±10 μA | 100 pA | 0.05% + 20 nA + Vo*500 pA | ||
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃, 28℃-50℃) | |||
I-V Out Capability

Typical Output Settling Time
| Source | Range | Output Settling Time[5] | Condition | ||
| Fast[6] | Normal | Slow | |||
| Voltage | 30 V | <500 μs | <1.5 ms | <3 ms | Time required to reach within 0.1 % of final value at open load condition. Step is 10 % to 90 % range |
| 6 V | <250 μs | <780 μs | <2.8 ms | ||
| Current | 500 mA | <150 μs | <400 μs | <2.5 ms | Time required to reach within 0.1 % of final value at short condition. Step is 10 % to 90 % range |
| 100 mA~1 mA | <120 μs | <300 μs | <2.5 ms | ||
| ±100 μA | <150 μs | <1.5 ms | <2.5 ms | ||
| ±10 μA | <700 μs | <2 ms | <2.5 ms | ||
[5] Output transition rate: Three modes of Fast, Normal, and Slow are available, allowing users to adjust according to load characteristics to achieve appropriate setup time or stability. This feature only supports switching the output transition rate when OUTPUT is set to OFF
[6] Slow mode is recommended for overshoot sensitive equipment. Fast mode may have overshoot on output in some condition