S2012C
S2012C
Single-channel PXIe SMU
Features

High Range and High-speed Measurement
± 200 V, ± 1 A (DC), ± 3 A (pulsed)
Up to 1M sampling rate can be supported

High Range and High-speed Measurement
± 200 V, ± 1 A (DC), ± 3 A (pulsed)
Up to 1M sampling rate can be supported

Adaptive PFC System
Leverages Adaptive PFC
(Precision Fast Control) system
Users can adjust relevant parameters according to load characteristics

Building Multi-channel Parallel Test System
Based on standard PXIe chassis
Easily implement multi-channel parallel testing
Functions and Advantages
5 Functions In One Card
Voltage source
Current source
Ammeter
Voltmeter
Electronic load


♦ The first and third quadrants are the source: the actual polarity of output V / I follows the source setting.
♦ The second and fourth quadrants are for load: CC and CV cooperate. When the load is used, the polarity of the load setting is opposite to the source polarity.
Can Test Various Equipment


Capture More Measurement Data
♦ 6.5-digit resolution: enjoy best-in-class 6.5-digit sourcing and measurement resolution.
♦ 10 fA / 100 nV resolution: excellent sensitivity for setting and measuring.
♦ 1M points / second: provide high-speed measurement, and can quickly set / digitize the rate to any waveform generator / list scan.
Rich Scanning Function


DC I-V Output Capacity
Pulse I-V Output Capacity

Voltage Programming and Measurement Specifications
| Voltage Accuracy | ||||
| Range | Programming Resolution | Accuracy (1 Year) ± (% reading+ offset)[1] |
Typical Noise(RMS) 0.1 Hz-10Hz |
|
| ±200 V[2] | 100 μV | 0.03%+10 mV | 0.4 mV | |
| ±20 V | 10 μV | 0.03%+1 mV | 50 μV | |
| ±6 V | 1 μV | 0.03%+0.4 mV | 12 μV | |
| ±0.6 V | 100 nV | 0.03%+100 μV | 3 μV | |
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃) | |||
| Overshoot | <±0.1% (typical.normal.step is 10% to 90% range, full range, resistive load) | |||
| Noise 10Hz-20MHz | <5 mVrms, (20V voltage source, 1A resistive load) | |||

[2] This instrument has a potentially dangerous high voltage (±210 V) output to the HI / Sense HI / Guard terminals. To prevent electric shock, relevant safety precautions must be taken before powering on. Do not connect the Guard terminal to any output, including shorting it to the chassis ground or output LO, as this will damage the instrument
Current Programming and Measurement Specifications
| Current Accuracy | ||||
| Range | Programming Resolution | Accuracy (1 Year)±(%reading+ offset) | Typical Noise(RMS) 0.1 Hz-10Hz |
|
| ±3 A[3] | 1 μA | 0.03% + 2mA | 40 μA | |
| ±1 A | 100 nA | 0.03% + 90 μA | 7 μA | |
| ±100 mA | 10 nA | 0.03% + 9 μA | 600 nA | |
| ±10 mA | 1 nA | 0.03% + 900 nA | 60 nA | |
| ±1 mA | 100 pA | 0.03% + 90 nA | 6 nA | |
| ±100 μA | 10 pA | 0.03% + 9 nA | 700 pA | |
| ±10 μA | 1 pA | 0.03% +1 nA | 80 pA | |
| ±1 μA[4] | 100 fA | 0.03% + 200 pA | 20 pA | |
| ±100 nA[4][5] | 100 fA | 0.06% + 30 pA | 3 pA | |
| ±10 nA[4][5] | 10 fA | 0.06% + 9 pA | 600 fA | |
| Temperature Coefficient | ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃) | |||
| Overshoot | <±0.1% (typical.normal.step is 10% to 90% range, full range, resistive load) | |||
[3] 3 A range is available only for pulse mode, with typical accuracy
[4] For accurate low current measurements, triaxial cable connection is recommended. Converting the triaxial output to standard wiring will affect current measurement accuracy
[5] Condition: NPLC = 10 PLC
Pulse Source Specifications
| Minimum Programmable Pulse Width | 100 μs | |||
| Pulse Width Programming Resolution | 1 μs | |||
| Pulse Width Programming Accuracy | ±10 μs | |||
| Pulse Width Jitter | 2 μs | |||
| Pulse Width Definition | The time from 10% leading to 90% trailing edge as follows | |||

| Maximums | Maximum Pulse Width | Maximum Duty Cycle | |
| 0.1 A/200 V | DC, no limit | 100% | |
| 1 A/20 V | DC, no limit | 100% | |
| 3 A/66.6 V | 1 ms | 5% | |
| 3 A/160 V | 400 μs | 2% | |
Typical Pulse Performance
| Source | Maximum Output | Typical Rise Time[6] | Typical Settling Time[7] | Test load |
| Voltage | 160 V | 800 μs | 1.2 ms | No load |
| 5 V | 80 μs | 200 μs | No load | |
| Current | 3A~1 mA | 120 μs | 250 μs | Full load[8] |
| 100 μA~10 μA | 200 μs | 400 μs | Full load[8] | |
| 1 μA | 800 μs | 2 ms | Full load[8] | |
| 100 nA | 2 ms | 5 ms | Full load[8] | |
| 10 nA | 5 ms | 20 ms | Full load[3] |
Sampling Rate and NPLC Setting
| Setting | Range | |||
| NPLC | 0.00005 PLC ~ 10 PLC | |||
| Sampling Rate | 5 sps ~ 1 Msps | |||