Single-channel PXIe SMU

The Semight S2012C is a compact and cost-effective, single-slot, single-channel PXIe Source/Measure Unit (SMU) with the capability to source and measure both voltage and current. It delivers output up to ±200 V, ±1 A (DC), and ±3 A (pulsed) with a constant power of 20 W and supports conventional SMU SCPI commands for easy test code migration. Compatible with existing mainstream PXIe chassis, it offers high integration, enabling easy channel expansion and multi-module synchronization. These features improve efficiency and lower the cost when integrating the SMUs into systems for production test.

Features

10

High Range and High-speed Measurement

± 200 V, ± 1 A (DC), ± 3 A (pulsed)
Up to 1M sampling rate can be supported

11

High Range and High-speed Measurement

± 200 V, ± 1 A (DC), ± 3 A (pulsed)
Up to 1M sampling rate can be supported

12

Adaptive PFC System

Leverages Adaptive PFC
(Precision Fast Control) system
Users can adjust relevant parameters according to load characteristics

13

Building Multi-channel Parallel Test System

Based on standard PXIe chassis
Easily implement multi-channel parallel testing

Functions and Advantages

5 Functions In One Card

Voltage source
Current source
Ammeter
Voltmeter
Electronic load

028cc969553b3060a5fcffc51afabf4d
Four Quadrants

♦ The first and third quadrants are the source: the actual polarity of output V / I follows the source setting.
♦ The second and fourth quadrants are for load: CC and CV cooperate. When the load is used, the polarity of the load setting is opposite to the source polarity.

Can Test Various Equipment

e7a81a57efa4cfce8ec085baeb0153e4
2ad4b20b5e7b6596405da5075554487c

Capture More Measurement Data

♦ 6.5-digit resolution: enjoy best-in-class 6.5-digit sourcing and measurement resolution.
♦ 10 fA / 100 nV resolution: excellent sensitivity for setting and measuring.
♦ 1M points / second: provide high-speed measurement, and can quickly set / digitize the rate to any waveform generator / list scan.

Rich Scanning Function

0749a0ac9b5011f3298c2b8431097577
S2012C DC I-V Output Capacity

DC I-V Output Capacity

Pulse I-V Output Capacity

S2012C Pulse I-V Output Capacity

Voltage Programming and Measurement Specifications

Voltage Accuracy
Range Programming Resolution Accuracy (1 Year)
± (% reading+ offset)[1]
Typical Noise(RMS)
0.1 Hz-10Hz
±200 V[2] 100 μV 0.03%+10 mV 0.4 mV
±20 V 10 μV 0.03%+1 mV 50 μV
±6 V 1 μV 0.03%+0.4 mV 12 μV
±0.6 V 100 nV 0.03%+100 μV 3 μV
Temperature Coefficient ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃)
Overshoot <±0.1% (typical.normal.step is 10% to 90% range, full range, resistive load)
Noise  10Hz-20MHz <5 mVrms, (20V voltage source, 1A resistive load)
[1] Accuracy calculation example: To test the accuracy of a 600mV range with a 120mV output, the tolerance is:241cfb3fae1b52c04f0a35c0075866f2
[2] This instrument has a potentially dangerous high voltage (±210 V) output to the HI / Sense HI / Guard terminals. To prevent electric shock, relevant safety precautions must be taken before powering on. Do not connect the Guard terminal to any output, including shorting it to the chassis ground or output LO, as this will damage the instrument

Current Programming and Measurement Specifications

Current Accuracy
Range Programming Resolution Accuracy (1 Year)±(%reading+ offset) Typical Noise(RMS)
0.1 Hz-10Hz
±3 A[3] 1 μA 0.03% + 2mA 40 μA
±1 A 100 nA 0.03% + 90 μA 7 μA
±100 mA 10 nA 0.03% + 9 μA 600 nA
±10 mA 1 nA 0.03% + 900 nA 60 nA
±1 mA 100 pA 0.03% + 90 nA 6 nA
±100 μA 10 pA 0.03% + 9 nA 700 pA
±10 μA 1 pA 0.03% +1 nA 80 pA
±1 μA[4] 100 fA 0.03% + 200 pA 20 pA
±100 nA[4][5] 100 fA 0.06% + 30 pA 3 pA
±10 nA[4][5] 10 fA 0.06% + 9 pA 600 fA
Temperature Coefficient ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃)
Overshoot <±0.1% (typical.normal.step is 10% to 90% range, full range, resistive load)

[3] 3 A range is available only for pulse mode, with typical accuracy
[4] For accurate low current measurements, triaxial cable connection is recommended. Converting the triaxial output to standard wiring will affect current measurement accuracy
[5] Condition: NPLC = 10 PLC

Pulse Source Specifications

Minimum Programmable Pulse Width 100 μs
Pulse Width Programming Resolution 1 μs
Pulse Width Programming Accuracy ±10 μs
Pulse Width Jitter 2 μs
Pulse Width Definition The time from 10% leading to 90% trailing edge as follows

438da186485e601a2e3b2c2278254381

Maximums Maximum Pulse Width Maximum Duty Cycle
0.1 A/200 V DC, no limit 100%
1 A/20 V DC, no limit 100%
3 A/66.6 V 1 ms 5%
3 A/160 V 400 μs 2%

Typical Pulse Performance

Source Maximum Output Typical Rise Time[6] Typical Settling Time[7] Test load
Voltage 160 V 800 μs 1.2 ms No load
5 V 80 μs 200 μs No load
Current 3A~1 mA 120 μs 250 μs Full load[8]
100 μA~10 μA 200 μs 400 μs Full load[8]
1 μA 800 μs 2 ms Full load[8]
100 nA 2 ms 5 ms Full load[8]
10 nA 5 ms 20 ms Full load[3]
[6] Time required for the pulse leading edge to rise from 10% to 90%
[7] Time required for the pulse to reach within 1% of final value
[8] Test conditions: Normal mode, resistive full load, voltage rises to 6 V

Sampling Rate and NPLC Setting

Setting Range
NPLC 0.00005 PLC ~ 10 PLC
Sampling Rate 5 sps ~ 1 Msps