S2019C
S2019C
4-channel PXIe SMU
The S2019C is a high precision, compact, cost-effective, 4-channel PXIe Source/Measure Unit (SMU) with the capability to source and measure both voltage and current. It has maximum ±40 V, ± 500 mA(DC), ±1 A(Pulse) sourcing capability, and supports conventional SCPI commands for easy programming. It supports most standard PXIe chassis and multi-card synchronization. The S2019C can be integrated into production test environment to achieve test efficiency improvement and cost reduction.
Features

Easy to Expand
4-channel PXIe SMU, easy to scale out

High Range
Range: ± 40 V, ± 500 mA (DC), ±1 A(pulse)

High Resolution
The minimum measurement resolution can reach 1 pA/1 μV

High Sampling Rate
Up to 1MS/s ADC sampling rate
Functions and Advantages
5 Functions In One Card
Voltage source
Current source
Ammeter
Voltmeter
Electronic load


♦ The first and third quadrants are the source: the actual polarity of output V / I follows the source setting.
♦ The second and fourth quadrants are for load: CC and CV cooperate. When the load is used, the polarity of the load setting is opposite to the source polarity.
Can Test Various Equipment


Capture More Measurement Data
♦ 6.5-digit resolution: Enjoy best-in-class 6.5-digit sourcing and measurement resolution.
♦ 1 pA / 1 μV resolution: excellent measurement sensitivity.
♦ 1 MS/s sampling rate: provide high-speed measurement, and can quickly set / digitize the rate to any waveform generator (list scan).
Rich Scanning Function

Voltage Source Setting and Measurement Accuracy
| Voltage Accuracy | ||||
| Range | Resolution | Accuracy(1 year)
±(%RD+mV)[1] |
Typical Noise (RMS)
0.1 Hz-10 Hz |
|
| ±40 V[3] | 10μV | 0.015%+1.2 mV | 50 μV | |
| ±6 V | 1 μV | 0.015%+600 μV | 10 μV | |
| Temperature Coefficient | ±(0.15×accuracy)/℃(0℃-18℃, 28℃-50℃) | |||
| Channel [2] | CH0 to CH3 | |||
| Output Power | 4 W per channel and maximum 8 W total | |||
| Overshoot | <±0.1%( typical, Normal, step is 10% to 90% range, full range, resistive load) | |||
| Noise
10Hz-20MHz |
6 V voltage source, 0.5 A resistive load, <3 mVrms | |||
[1] Accuracy calculation example: Testing the accuracy of a 1V output in a 6V range, the allowable tolerance is:![]()
[2] All channel outputs are electrically isolated from the ground, but the outputs of each channel share a common ground(LO)
[3] This instrument has a potential danger of high voltage (±42 V) output to the HI/Sense HI/Guard terminals. To prevent electric shock, relevant safety precautions must be taken before powering on. Do not connect the Guard terminal to any output. (Including shorting to the chassis ground or outputLO),otherwise, the instrument will be damaged
Current Source Setting and Measurement Accuracy
| Current
Accuracy |
||||
| Range | Resolution | Accuracy(1 year) ±(%RD+mA) |
Typical Noise (RMS)
0.1Hz-10Hz |
|
| ±1 A[4][5] | 500 nA | 0.1% + 125 μA | 20 μA | |
| ±500 mA[5] | ||||
| ±100 mA | 100 nA | 0.03% + 25 μA | 2 μA | |
| ±10 mA | 10 nA | 0.03% + 2.5 μA | 200 nA | |
| ±1 mA | 1 nA | 0.03% + 250 nA | 20 nA | |
| ±100 μA | 100 pA | 0.03% + 25 nA | 2 nA | |
| ±10 μA | 10 pA | 0.03% + 3 nA | 200 pA | |
| ±1 μA | 1 pA | 0.03% + 500 pA | 30 pA | |
| Temperature Coefficient | ±(0.15×accuracy)/℃(0℃-18℃,28℃-50℃) | |||
| Channel | CH0 to CH3 | |||
| Output Power | 4 W per channel and maximum 8 W total | |||
| Overshoot | <±0.1%(typical. normal. step is 10% to 90% range, full range, resistive load) | |||
[4] The 1A range only supports pulse mode, with a typical accuracy
[5] The 1A and 500mA ranges only support a voltage range of 6V