S2022H

S2022H

8-channel High-precision SMU

The S2022H is a high-precision, compact and cost-effective benchtop Source/ Measure Unit (SMU) with the capability to source and measure both voltage and current. The S2022H is ideal for IV (current versus voltage) measurement tasks that require both high resolution and accuracy.

Features

10

High Range

Range: ± 30 V, ± 500 mA (DC/pulse)

11

High Resolution

The minimum measurement resolution can reach 100 pA/60 μV

12

High Rampling Rate

Supports up to 500K ADC sampling rate

13

Threshold Trigger

Hardware high-speed IO, capable of threshold triggering, enabling efficient interaction between output measurement values and user systems

Functions and Advantages

I-V Output Capability

I V Output Capability 2

Voltage Source Specifications

Voltage Programming Accuracy
Range Programming Resolution Accuracy (1 year)

± (% reading+offset)[1]

Typical Noise (RMS)

0.1 Hz-10 Hz

± 30 V[3] 1  mV 0.03%+4 mV 1000 μV
± 6 V 200 μV 0.03%+1 mV 100 μV
Temperature Coefficient ± (0.15 x accuracy)/℃ (0 ℃ -18 ℃, 28 ℃ -50 ℃)
Channel[2] CH1 to CH8
Output Power Ch1 Ch4: 3 W per channel and 6W total

Ch5 Ch8: 3 W per channel and 6W total

Overshoot < ± 0.1% (typical, normal, step is 10% to 90%, full range, resistive load)
Noise 10Hz-20MHz 6V voltage source, 0.5A resistive load,< 3 mVrms

[1] Accuracy calculation example: For testing the accuracy of a 6V range with a 1V output, the tolerance is

084fcd02c92a2e84700df2ffe5eb4179

[2] All channel outputs are electrically isolated from ground, but the outputs of channels CH1-CH4 share a common ground (LO)
[3] This instrument has potentially dangerous high voltage (±31.5 V) output to the HI / Sense HI / Guard terminals. To prevent electric shock, relevant precautions must be taken before powering on. Safety precautions. Do not connect the Guard terminal to any output, including shorting it to chassis ground or output LO, as this will damage the instrument.

Current Source Specifications

Current Programming Accuracy
Range Programming  Resolution Accuracy (1 year)

± (% reading+offset)

Typical Noise (RMS)

0.1 Hz-10 Hz

± 500 mA[4] 20 μA 0.05%+100 μA+  Vo * 25 μA 10 μA
± 100 mA 4  μA 0.05%+10 μA+  Vo * 5 μA 1 μA
± 10 mA 400 nA 0.05%+5 μA+  Vo * 500 nA 100 nA
± 1 mA 40 nA 0.05%+500 μA+ Vo * 50 nA 10 nA
± 100 μA 4  nA 0.05%+50 nA+  Vo * 5 nA 1 nA
± 10 μA 400 pA 0.05%+20 nA+  Vo * 500 pA 150 pA
Temperature Coefficient ± (0.15 x accuracy)/℃ (0 ℃ -18 ℃, 28 ℃ -50 ℃)
Channel[5] CH1 to CH8
Output Power Ch1 Ch4: 3W per channel and 6W total

Ch5 Ch8: 3W per channel and 6W total

Overcharge < ± 0.1% (typical, normal, step is 10% to 90%, full scale point, resistive load test)

[4] 500mA range is available only for 6V voltage range
[5] Channels are isolated from earth ground, but CH1 to CH4 share a common LO, and CH5 to CH8 share a common LO

Voltage Measurement Specificationgs

Voltage Measurement Accuracy
Range Measurement Resolution Accuracy (1 year)

± (% reading+offset)

± 30 V 300 μV 0.03%+4 mV
± 6 V 60 μV 0.03%+1 mV
Temperature Coefficient ± (0.15 x accuracy)/℃ (0 ℃ -18 ℃, 28 ℃ -50 ℃)

Current Measurement Specifications

Current Measurement Accuracy
Range Measurement Resolution Accuracy (1 year)

± (% reading+offset)

± 500 mA[6] 10 μA 0.05%+100 μA+  Vo * 25 μA
± 100 mA 1 μA 0.05%+10 μA+  Vo * 5 μA
± 10 mA 100 nA 0.05%+5 μA+  Vo * 500 nA
± 1 mA 10 nA 0.05%+500 nA+  Vo * 50 nA
± 100 μA 1 nA 0.05%+50 nA+  Vo * 5 nA
± 10 μA 100 pA 0.05%+20 nA+  Vo * 500 pA
Temperature Coefficient ± (0.15 x accuracy)/℃ (0 ℃ -18 ℃, 28 ℃ -50 ℃)

[6] 500mA range is available only for 6V voltage range

Typical Output Setting Time

Source Range Output setting time[7] Condition
Fast [8] Normal Slow
Voltage 30 V < 400 μs < 1.5ms < 2.8 ms Time required to reach within 0.1 % of final value at open load condition.
Step is 10 % to 90 % range
6  V <250 μs <780 μs <2.8 ms
Current ± 500 mA <50 μs <330 μs <2.5 ms Time required to reach within 0.1 % of final value at short condition.
Step is 10 % to 90 % range
±100 mA <50 μs <270 μs <2.5 ms
±10 mA <50 μs <270 μs <2.5 ms
±1 mA <100 μs <290 μs <2.5 ms
±100 μA <150 μs <5 ms <2.5 ms
±10 μA <250 μs <3 ms <2.5 ms

[7] Output transition rate: Three modes of Fast, Normal, and Slow are available, allowing users to adjust according to load characteristics to achieve appropriate setup time or stability. This feature only supports switching the output transition rate when OUTPUT is set to OFF
[8] The Fast mode may produce a large overshoot under different ranges or load conditions. For overshoot-sensitive devices, it is recommended to use normal or Slow mode. mode