S3029H

S3029H

Precision Pulse SMU

S3029h is a single channel, pulse current source aimed for testing LiDAR. It supports high-precision pulse current output and synchronous measurement of pulse current with DUT voltage drop.

Features

10
Pulse Width

Minimum 3 μs, maximum 500 μs
Suitable for faster tested devices to better avoid the impact of temperature on test results

11
High-speed Measurement

ADC up to 100 MSa/s

12
High Range

Range: ±1.5 A, 10 V

13
Stronger Load Adaptability

Test loads with different characteristics to ensure that the waveform does not experience overshoot or other distortions

17
Closed-loop Current Control

Higher current accuracy

15
Real Time Reading of DUT Voltage Drop for Pulse Current

No additional equipment required, precise monitoring of current testing conditions, real-time reading back of test results

Functions and Advantages

  • Working conditions
  • Temperature 23 ℃± 5 ℃
  • Relative humidity<70%

Pulse Current Source (exclusive with Voltage Source)

Parameter
Specification 
Current Accuracy Range Programming Resolution  Accuracy

±(%RD+mA)

±150mA

20μA 0.1%+0.2mA
±750mA 20μA 0.1%+0.3mA
±1.5A 40μA 0.1%+0.8mA
Current Noise Range Noise Typical Value (Resistance Load,RMS)

10k-20MHz

 

±150mA

1mA

 

±750mA 1mA

 

±1.5A 1mA
Maximum Load Voltage Drop 10V
Pulse Width Resolution 80ns 
Maximum Pulse Width Ton-max 500μs
Minimum Pulse Width Ton-min 3μs
Pulse Minimum Turn-off Time Toff-min 100μs
Pulse Width Accuracy 100ns[1]
Pulse Width Jitter 80ns(typical value)
Pulse Period Jitter 500ns(typical value)
Rising Time(10%-90%) <400ns[2][3][4] 
Pulse Overshoot <0.5%[2][3][4] 
Current Regulation Rate

Linear 0.05% of Range
 
Load ±100μA
 
Duty Cycle[5] D≤50% when meet:

D<3-|Ibias|/[(Vsp-Vload)*(|Iset|-|Ibias|)]

Maximum Number of Pulses per Scan 64K

 

[1] Bias current>1mA

[2] Test with short of output line

[3] Total inductance of output line and DUT less than<200nH(100k)

[4] Pulse width measured from 10% of rising edge to 90% of falling edge

[5] Iset:programming current; D: duty cycle;Vsp:source protection voltage; Vload:load voltage

Pulse Measurement

Parameter Specification
Pulse

Voltage

Measurement

Range Display Resolution Accuracy

±(%RD+mV)

Sampling Rate [6]
6V 1mV 0.1%+22mV 100MSa/s
10V 1.5mV 0.1%+24mV
Pulse

Current

Measurement

Range Display Resolution Accuracy

±(%RD+mA)

Sampling Rate
150mA 40μA 0.1%+1mA 100MSa/s
750mA 100μA 0.1%+3mA
1.5A 200μA 0.1%+7mA
Remote Sense Limit (Voltage Measure)

Maximum voltage between HI and SENSE HI = ±10V;

Maximum voltage between LO and SENSE LO = ±10V;

 

[6] Maximal sampling rate for user setting

DC Voltage Source (exclusive with Pulse Current Source)

Parameter Specification
Voltage Source(DC) Range Programming Resolution Accuracy

±(%RD+mV)

Ripple(RMS)

10k-20MHz

Maximum Output Current
±10V 1mV 0.1%+1.5mV <5.5mV 10mA
Current Measure(DC) Range Programming Resolution Accuracy

±(%RD+nA)

Sampling Rate NPLC Setting
100μA 30nA 0.1%+0.5μA 100MSa/s  

 

10 NPLC

100nA 30pA 0.1%+0.5nA
Voltage Measure(DC) Range Programming Resolution Accuracy

±(%RD+mV)

Sampling Rate
±10V 1.5mV 0.1%+15mV 100MSa/s
Remote Sense Limit (DC Voltage Measure) Maximum voltage between HI and SENSE HI = ±1V;

Maximum voltage between LO and SENSE LO = ±1V;

Bias Current Source

Parameter

Specifications
Current

Accuracy

Range Maximum Current Programming Resolution Accuracy
±(%RD+mA)
±0.15A 50mA 20μA 0.1%+0.2mA
±0.75A 20μA 0.1%+0.2mA
±1.5A 40μA 0.1%+0.2mA
Current Noise Noise Typical Value(Resistive Load,RMS)

10k-20MHz

<1mA