S2014C
S2014C
12-channel PXIe SMU
The S2014C is a high precision, compact, cost-effective, 12-channel PXIe Source/Measure Unit (SMU) with the capability to source and measure both voltage and current. It has maximum ±10 V, ± 50mA (DC/pulse) sourcing capability. It supports conventional SCPI commands for easy programming. It supports most standard PXIe chassis and multi-card synchronization. The S2014C can be integrated into production test environment to achieve test efficiency improvement and cost reduction.
Features

High range, High-speed Measurement
±10 V, ±50 mA(DC/pulse)
Resolution up to 1pA/1μV
Supports up to 1M sampling rate

12-channel Testing System
Based on standard PXIe chassis
enabling multi-channel expansion and seamless
integration into rack-mounted and stacked systems

Adaptive PFC System
Utilizing Adaptive System
( Adaptive Precise-Fast Control)
Users can adjust relevant parameters based on load characteristics

Sensing Mode
Supports 2-wire or 4-wire (remote sensing) connection
Maximum sensor lead resistance: 1kΩ (rated accuracy)
Maximum voltage of remote output terminal and sensor terminal: 0.5V
Functions and Advantages
5 Functions In One Card
- Voltage sour
- Current source
- Ammeter
- Voltmeter
- Electronic load


♦ The first and third quadrants are the source: the actual polarity of the output V/I follows the source setting.
♦ The second ang fourth quadrants are for load: CC and CV cooperate. When the load is used, the polarity of the load setting is opposite to the source polarity.
Test Various Devices


Capture More Measurement Data
Rich Scanning Functions

Voltage Programming and Measurement Specifications
| Voltage Precision
|
||||
| Measuring Range[1] | Measuring Resolution | Accuracy (1 year)
± (% reading + bias)[2] |
Typical Noise (RMS)
0.1 Hz-10Hz |
|
| ±10V | 1 μV | 0.02%+500μV | 100 μV | |
| ±6V | 1 μV | 0.02%+300μV | 50 μV | |
| Temperature Coefficient | ±(0.15 × Precision Indicator)/°C (0℃-18℃,28℃-50℃) | |||
| Channel[3] | CH1-CH12 | |||
| Overshoot | <±0.1% (typical,normal, step range is 10% to 90%,full scale point,resistive load test) | |||
| Noise(10Hz-20MHz) | <3 mVrms,10V voltage source,50 mA resistive load | |||
[1] Please do not connect the Guard terminal to any output, including shorting to chassis ground or output LO, as it may damage the instrument
[2] Example of accuracy calculation: Testing the accuracy of a 1V output in a 10V range, the allowable tolerance is

[3] All channel outputs are electrically isolated from earth, but each channel output shares a common ground (LO)
Current Programming and Measurement Specifications
| Current Precision
|
||||
| Measuring Range | Measuring Resolution | Accuracy (1 year)
± (% reading + bias) |
Typical Noise (RMS)
0.1 Hz-10Hz |
|
| ±50 mA | 10 nA | 0.05%+12μA | 1 μA | |
| ±10 mA | 10 nA | 0.05%+2 μA | 100 nA | |
| ±1 mA | 1 nA | 0.05%+200nA | 10 nA | |
| ±100 μA | 100 pA | 0.05%+20nA | 1 nA | |
| ±10 μA | 10 pA | 0.05%+2nA | 100 pA | |
| ±1 μA | 1 pA | 0.05%+300pA | 50 pA | |
| Temperature Coefficient | ±(0.15 × precision indicator)/°C (0℃-18℃,28℃-50℃) | |||
| Channel | CH1-CH12 | |||
| Overshoot | <±0.1% (typical value,normal, step range is 10% to 90%,full scale point,resistive load test) | |||
Sampling Rate and NPLC Setting
| Configuration Method | Configuration Range |
| NPLC | 0.00005PLC ~ 10PLC |
| Sampling Rate | 5sps ~ 1.0Msps |