S3023P
High-Voltage Semiconductor Pulse Generator Unit S3023P
Semight S3023P It is a Single-Channel High-Voltage Semiconductor Pulse Generator Unit that can output adjustable high-precision high-voltage pulse sources.
- High voltage range – ±40V,Meet NVM testing technical conditions
- High voltage with fast rate of change – 1000V/us,A high-voltage pulse source that can generate faster edges
- Standard pulse mode – three-level pulse, period:40ns~10s
- Delay adjustable – time:0~9.999s,resolution ratio:2.5ns
- High-speed measurement – 500MS/s,Timely diagnosis of load impedance
- Adjustable pulse count – range:1~100,000个
- Minimum sampling period for voltage detection – 5us
- Level timing trigger – TTLLevel, pulse cycle synchronization
Features and Benefits
Working conditions:
Temperature 23 ℃± 5 ℃
Humidity 30% to 70% relative humidity measure after preheating for 60 minutes, and the ambient temperature change during measurement < ± 3 ° C
Calibration cycle is 1 year
Measurement speed 1 PLC

Pulse or DC output voltage indicators
Project | Test conditions | Technical indicators | Remarks |
Voltage amplitude | 50 Ω load | -20V~+20V | |
Open circuit | -40V~+40V | ||
Voltage amplitude resolution | 50 Ω load | 0.2mV | |Vout | ≤ 5V |
0.8mV | 0.5V ≤ | Vout | ≤ 20V | ||
open circuit | 0.4mV | | Vout | ≤ 10V | |
1.6mV | 10V ≤ | Vout | ≤ 40V | ||
Output interface | SMA | ||
Internal resistance | 50 Ω ± 1% | ||
Output short-circuit current | 800mA peak (400mA mean) | ||
Output overshoot (ringing) | 50 Ω load | ± (5%+20mV) |
Pulse or DC output voltage indicators (both with 50 Ω terminal load, should be open-loop control indicators, high-speed pulse)
Project | Sub projects | Technical indicators | Remarks |
Frequency range | 0.1~33MHz | ||
Pulse period | Programmable range | 40ns~10s | |
Resolution | 10ns | ||
Minimum cycle | 100ns | Vamp ≤ 10V | |
Accuracy | ± 1% (0.01%+200ps) | ||
Pulse width | Programmable range | 20ns~(pulse period -10ns) | |
Resolution | 2.5ns | Vamp< 10V | |
20ns | Vamp>10V | ||
Minimum value (10%~90% edge) | 120ns | ||
Accuracy | ± (3%+2ns) | ||
Pulse process time (Tr and Tf) | Programmable range | 20ns~400ms | |
Resolution | 2ns | Tr+Tf ≤ 10us | |
20ns | Tr+Tf>10us | ||
Minimum value (10%~90% edge) | 30ns | Vamp ≤ 5V | |
35ns | Vamp ≤ 10V | ||
40ns | Vamp ≤ 20V | ||
60ns | Vamp>20V | ||
Accuracy | -5%~+5%+25ns | Vamp ≤ 10V | |
-5%~+5%+40ns | Vamp ≤ 20V |
SPGU supplementary indicators
Advanced features:
Project | Technical indicators | Test conditions |
Pulse width jitter | 0.001%+150ps | |
Pulse period jitter | 0.001%+150ps | |
Maximum voltage swing rate | 1000V/us | Vamp=40V (50 Ω Load) |
Output noise | 20mVrms | |
Voltage monitoring | SPGU monitoring and testing unit terminal voltage | |
Measurement accuracy | ± (0.1%+25mV) | |
Measurement resolution | 50uV |