S3030F

Single-Channel High voltage Precision Source Meter

The Semight S3030F is a compact and cost-effective single channel high voltage, high-power power supply/measurement unit that can simultaneously output and measure voltage and current, it can provide a maximum power output of ± 3500V, ± 120mA (DC), 10 G Ω, and 180W, and can be widely used in testing and research fields such as power semiconductor characteristics, GaN, SiC characterization, composite materials, and high-voltage leakage current. S3030F supports traditional SMU SCPI commands, making migration of test code easy and fast, it can support multi machine synchronization and be integrated into production testing systems to improve system testing efficiency and reduce costs.

  • High range – Range:±3500 V、±120 mA(DC)、180 W
  • High resolution – The minimum measurement resolution can reach 10 fA/100 μV
  • High sampling rate – Supports up to 1M ADC sampling rate
  • Threshold trigger – Hardware high-speed IO, capable of threshold triggering, enabling efficient interaction between output measurement values and user systems

Features and Benefits

DC I-V output capability

working conditions:
Temperature 23 °C ± 5 °C
Humidity 30% to 70% relative humidity
Measure after preheating for 60 minutes, and the ambient temperature change during measurement is less than ± 3 ° C
Calibration cycle 1 year
Measurement speed 1 PLC

910607b9e6c08d3502812cea42b34690 png

Voltage Source specifications

Voltage setting accuracy Range Programming resolution Accuracy (1 year)

± (% reading+offset)

Typical noise(RMS)

0.1Hz-10Hz

±3500 V 40 mV 0.02%+600 mV 50 mV
±2500 V 30 mV 0.02%+450 mV 40 mV
±1500 V 20 mV 0.02%+300 mV 25 mV
±600 V 7 mV 0.02%+120 mV 10 mV
±200 V 3 mV 0.02%+40 mV 3 mV
Temperature Coefficient ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃)
Settling time <5 ms (typical)
Overshoot <± 1 % (Typical. Norma mode. Step is 10 % to 90 % range, full range, resistive load)
Noise 10Hz-20MHz 1500 V voltage source, 120 mA resistive load, < 200 mV RMS

Current Source specifications

Current setting accuracy Range Programming resolution Accuracy (1 year)

± (% reading+offest)

Typical noise (RMS)

0.1Hz-10Hz

±120 mA1 3 μA 0.02% + 35 μA 120 uA
±20 mA 400 nA 0.02% + 15 μA 20 uA
±10 mA 200 nA 0.02% + 3 uA 10 uA
±1 mA 20 nA 0.02% + 300 nA 1 uA
±100 μA 2 nA 0.02% + 30 nA 100 nA
±10 μA 200 pA 0.03% + 5 nA 10 nA
±1 μA 20 pA 0.03% + 1 nA 1 nA
±100 nA 2 pA 0.2% +100 pA+Vo x 100 fA 100 pA
±10 nA 200 fA 0.2% + 10 pA+ Vo x 10 fA 10 pA
±1 nA 20 fA 0.2% + 5 pA+ Vo x 1 fA 1 pA
Temperature Coefficient ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃)
Settling time <10 ms (typical)
Overshoot <± 1 % (Typical. Norma mode. Step is 10 % to 90 % range, full range, resistive

load)

 

  1. 120mA range only supports 1500V and below voltage range

Voltage measurement specifications

Voltage measurement accuracy

 

 

 

 

 

Range Measurement resolution Accuracy (1 year)

± (% reading+offset)

±3500 V 1 mV 0.02%+600 mV
±2500 V 1 mV 0.02%+450 mV
±1500 V 1 mV 0.02%+300 mV
±600 V 100 μV 0.02%+120 mV
±200 V 100 μV 0.02%+40 mV
Temperature Coefficient ±(0.15 × accuracy)/°C (0℃-18℃,28℃-50℃)

Current measurement specifications

Current measurement accuracy

 

 

 

 

 

 

 

 

 

Range Measurement resolution Accuracy (1 year)

± (% reading+oddset)

±120 mA1 100 nA 0.02% + 35 μA
±20 mA 10 nA 0.02% + 15 μA
±10 mA 10 nA 0.02% + 3 μA
±1 mA 1 nA 0.02% + 300 nA
±100 μA 100 pA 0.02% + 30 nA
±10 μA 10 pA 0.03% + 5 nA
±1 μA 1 pA 0.03% + 1 nA
±100 nA 100 fA 0.2% +100 pA+Vo x 100 fA
±10 nA 10 fA 0.2% + 10 pA+ Vo x 10 fA
±1 nA 1 fA 0.2% + 5 pA+ Vo x 1 fA
Temperature

Coefficient

±(0.15 × accuracy index)/°C (0℃-18℃,28℃-50℃)

 

  1. 120mA range only supports 1500V and below voltage range.