S2024H
24-Channel Precision Source Meter
The Semight S2024H is a compact and cost-effective 24 channel desktop power supply/measurement unit (SMU) that can simultaneously output and measure voltage and current, providing maximum ± 4.5V, ± 10 mA (DC/pulse) output, and excellent color LCD graphical user interface (GUI).
- High range – Range: ± 4.5 V, ± 10 mA (DC)
- High resolution – The minimum measurement resolution can reach 10 pA/1 μV
- High sampling rate – Supports up to 1M ADC sampling rate
- Threshold trigger – Hardware high-speed IO, capable of threshold triggering, enabling efficient interaction between output measurement values and user systems
Voltage source/ measurement specifications
Voltage programming accuracy | Range | Measurement resolution | Accuracy (1 Year)
± (% reading+ offset) |
Typical Noise (RMS)
0.1 Hz-10 Hz |
±4.5V | 1 μV | 0.02%+100 μV | 50 uV | |
Temperature coefficient | ±(0.15 × accuracy)/℃ (0℃-18℃, 28℃-50℃) | |||
Settling time | <100 μs (typical) | |||
Overshoot | <±0.1% (Typical. Norma mode. Step is 10 % to 90 % range, full range, resistive load) |
Current source/ measurement specifications
Current programming accuracy | Range | Measurement resolution | Accuracy (1 Year) ± (% reading+ offset) | Typical Noise (RMS) 0.1 Hz-10 Hz |
±10 mA | 10 nA | 0.05%+5 μA | 20 nA | |
±1 mA | 1 nA | 0.05%+500 nA | 10 nA | |
±100 μA | 100 pA | 0.05%+50 nA | 1 nA | |
±10 μA | 10 pA | 0.05%+5 nA | 100 pA | |
Temperature coefficient | ±(0.15 × accuracy)/℃ (0℃-18℃,28℃-50℃) | |||
Settling time | <2ms (typical) | |||
Overshoot | <±0.1% (Typical. Normal mode. Step is 10 % to 90 % range, full range, resistive load) |
- Channels are isolated from earth ground but share a common LO.
Pulse source specifications (4W)
Minimum programmable pulse width | 250 μs |
Pulse width programming resolution | 1 μs |
Pulse width programming accuracy | ±10 μs |
Pulse width jitter | 2 μs |
Pulse width definition | The time from 10 % leading to 90 % trailing edge as follows |
Typical Pulse Performance(4W)
Source | Maximum output | Typical rise time1 | Typical Settling Time2 | Test load |
Voltage | 4.5 V | 100 μs | 200 μs | NO load |
Current | 10 mA | 60 μs | 100 μs | Full load3 |
1 mA | 800 μs | 1 ms | Full load3 | |
100 uA | 120 μs | 180 μs | Full load3 | |
10 μA | 1.5 ms | 2 ms | Full load3 |
- Leading edge, the time from 10 % leading to 90 % leading
- The time required from Pulse out 0 to reach within 1 % of final value
- Test condition: Normal, resistive load 4.5V maximum output
Sampling rate and NPLC setting
Setting | Range |
NPLC | 0.00005PLC ~ 10PLC |
Sampling Rate | 5sps ~ 1.0Msps |
Derating accuracy with PLC setting< 1 PLC
Add % of range using the following table for measurement with PLC < 1
PLC | RANGE | ||||
4.5 V | 10 μA | 100 μA | 1 mA | 10 mA | |
0.1 | 0.01% | 0.02% | 0.01% | 0.02% | 0.01% |
0.01 | 0.3% | 0.2% | 0.04% | 0.04% | 0.02% |
0.001 | 3.2% | 2.5% | 0.4% | 0.3% | 0.03% |